Page 1 - Medical Weight Analyzer

Model MV1 Medical Weight Analyzer Owner’s Manual 8555-M512-O1 Rev A 03/13 1

Page 2

Mounting The MedVue Model MV1 Medical Weight Analyzer is mounted to the scale (or an optional desktop/wall bracket) using a swivel ball and socket ass

Page 3 - PRECAUTIONS

INTERCONNECTIONS All input, output, power and option connections to the MV1 are made on the its rear panel. Connections for the Load Cell input and t

Page 4 - CAUTION

Serial I/O Port The serial I/O port is a 9-pin "D" connector on the rear panel of the MV1. Figure No. 5 shows the Serial I/0 connector alon

Page 5 - TABLE OF CONTENTS

Battery Status Symbol A battery status symbol will be displayed when batteries are installed. The number of bars indicates the level of charge remain

Page 6

KEYPAD FUNCTIONS DO NOT operate the keypad with pointed objects (pencils, pens, etc). Damage to keypad resulting from this practice is NOT covered u

Page 7 - SPECIFICATIONS

Basic Functions, Cont. This is the LOCK/RELEASE key. It is used to lock and unlock the display. If the HOLD feature was enabled during setup and ca

Page 8 - Electrical Power

Body Mass Index (BMI) Functions This is the CLEAR (BACKSPACE) key. It is used to clear an incorrect entry from the display without processing the da

Page 9 - INSTALLATION

ANNUNCIATORS The annunciators are turned on to indicate that the display is in the mode corresponding to the annunciator label or that the status indi

Page 10 - Mounting

CAL The CAL annunciator will be turned on after pressing the calibration switch, to show that the MV1 is in the Setup and Calibration mode. lb This a

Page 11 - 8555-M512-O1 Rev A 03/13

OPERATION Basic Operation To Weigh 1. Press ON/OFF key to turn the MV1 on. 2. Press ZERO key to zero weight display. ZERO, GROSS and lb or kg annunci

Page 12 - BATTERIES

8555-M512-O1 Rev A 03/13 2

Page 13 - Battery Status Symbol

Tare Weight Entry (Tare weight is known) 1. In Gross Weight mode (GROSS annunciator on), press the TARE key. 2. The Display will change to show ENT

Page 14 - KEYPAD FUNCTIONS

Tare Storage and Recall NOTE: ID must be enabled to use stored Tare. Refer to SYSTEM Menu, Enable ID. To Store Tare Weight 1. In Gross Weight mod

Page 15 - Basic Functions, Cont

Basic Operation with ID To Weigh 1. Press ON/OFF key to turn the MV1 on. 2. Press ZERO key to zero weight display. ZERO, GROSS and lb or kg annuncia

Page 16

Tare Weight Entry (Tare weight is known) 1. In Gross Weight mode (GROSS annunciator on), press TARE key. 2. Display will show )0 or previously ente

Page 17 - ANNUNCIATORS

Body Mass Index (BMI) Operation Using Mechanical Height Rod or Manually Entering Height 1. Press ON/OFF key to turn the MV1 on. 2. Press ZERO key to

Page 18

Body Mass Index (BMI) Operation with DHR Using Digital Height Rod IMPORTANT! Before performing the operations on this page, the digital height rod m

Page 19 - OPERATION

Body Mass Index (BMI) Operation with ID and DHR Using Digital Height Rod IMPORTANT! Before performing the operations on this page, the digital heigh

Page 20

CALIBRATION SETUP (Protected Setup Parameters) Your MedVue Model MV1 Medical Weight Analyzer has been thoroughly tested and calibrated before being s

Page 21 - Tare Storage and Recall

Navigation Keys, Cont. ID/HEIGHT – (Next Menu or Parameter) If at the menu level, this will take you to the previous menu item in the list below. If

Page 22 - Basic Operation with ID

SCALE With SCALE displayed, press the PRINT/ENTER key. The display will change to CAPACITY. Proceed to the CAPACITY (Scale Capacity) parameter. O

Page 23

INTRODUCTION Thank you for purchasing our MedVue® Model MV1 Medical Weight Analyzer. It has been manufactured with quality and reliability at our fac

Page 24

SAMPLE RATE (Sample Rate) The display will show SAMPLE RATE and the current setting. If the value displayed is acceptable, press the PRINT/ENTER key

Page 25 - 110.0 cm

ZERO With ZERO displayed, press the PRINT/ENTER key. The display will change to ZERO LIMIT. Proceed to the Two Percent Zero Tracking Range Limit par

Page 26

FILTER With FILTER displayed, press the PRINT/ENTER key. The display will change to FILTER MOdE. Proceed to the Filter Type parameter. Otherwise, p

Page 27 - CALIBRATION SETUP

CALIbRATION With CALIbRATION displayed, press the PRINT/ENTER key. The display will change to CALIbRATE. Proceed to the Perform Calibration paramet

Page 28

Single Calibration Point LOAdEd CAL WT (Load Calibration Weight) The display will show LOAdEd CAL WT (which is a prompt for entry of the calibration w

Page 29

Multi-point Calibration CAL WEIGHT 1 (First Calibration Weight) The display will show CAL WEIGHT 1 (which is a prompt for entry of the first calibrati

Page 30 - SAMPLE RATE (Sample Rate)

GRAVITY With GRAVITY displayed, press the PRINT/ENTER key. The display will change to GRAVITY ENAbLE. Proceed to the Enable Gravity Constant param

Page 31

LOCK With LOCK displayed, press the PRINT/ENTER key. The display will change to HOLd MOdE. Proceed to the Hold Mode parameter. Otherwise, press the

Page 32 - FILTER MOdE (Filter Type)

8555-M512-O1 Rev A 03/13 32

Page 33 - CALIbRATION

CALIBRATION SETUP (Non- Protected Setup Parameters) SYSTEM With SYSTEM displayed, press the PRINT/ENTER key. The display will change to LANGUAGE. Pr

Page 34 - Single Calibration Point

FCC COMPLIANCE STATEMENT This equipment generates uses and can radiate radio frequency and if not installed and used in accordance with the instructio

Page 35 - Multi-point Calibration

SLEEP MOdE (Sleep Mode) The Sleep Mode feature also conserves power when the MV1 remains unused for a selected period of time (approximately 1 to 9 mi

Page 36 - GRAVITY

bARCOdE (Use ID Scanner) This setting determines whether the MV1 allows a scanner to be used to input an ID number to be entered prior to beginning th

Page 37 - HOLd MOdE (Hold Mode)

HI RESOLUTION (High Resolution Mode) The High Resolution Mode feature will display the weight at 10 times the standard resolution. For example, a MV1

Page 38

SERIAL PORT With SERIAL PORT displayed, press the PRINT/ENTER key. The display will change to RS232 MOdE. Proceed to the RS232 Port Output Mode para

Page 39

USb PORT With USb PORT displayed, press the PRINT/ENTER key. The display will change to USb MOdE. Proceed to the USB Port Output Mode parameter. Ot

Page 40 - SLEEP MOdE (Sleep Mode)

ETHERNET/WIFI – (If Option Card Installed) With ETHERNET/WIFI displayed, press the PRINT/ENTER key. The display will change to ETHERNET MOdE. Procee

Page 41

TIME SETTINGS With TIME SETTINGS displayed, press the PRINT/ENTER key. The display will change to YEAR [YYYY]. Proceed to the YEAR parameter. Other

Page 42

24 HOUR TIME (Use 24 Hour Time Format) The display will show 24 HOUR TIME and the current setting. If the value displayed is acceptable, press the PR

Page 43 - SERIAL PORT

8555-M512-O1 Rev A 03/13 42

Page 44 - USb PORT

SETUP REVIEW (Non-Protected Setup Parameters) The MedVue Model MV1 Medical Weight Analyzer allows several operational parameters to be reviewed and ch

Page 45 - If Option Card Installed)

TABLE OF CONTENTS SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 1 Standard Fea

Page 46 - TIME SETTINGS

SERIAL PORT RS232 MOdE Set the mode of the RS232 serial port bAUd RATE Set baud rate for RS232 serial port (If RS232 MOdE = 1, 2 or 3) PRINTER TYPE

Page 47

ERROR AND STATUS DISPLAYS Display Meaning ERROR General error, invalid keypad entry was attempted. ERR OVERFLOW Attempting to display a negative numb

Page 48

BEFORE YOU CALL FOR SERVICE Problem Possible Solutions Display does not turn on AC Operation:  Is AC power supply fully inserted into wall recepta

Page 49 - SETUP REVIEW

SECURITY SEAL INSTALLATION If your metrology laws require the use of physical sealing, a lead and wire security seal can be installed on the MV1 to pr

Page 50

PARTS IDENTIFICATION Item Qty Part Number Description 1 1 593GR986 SERIAL TAG ASSY 2 1 6007-0009 BOLT HEX HD 3 1 6021-0423 SCW PAN-HEAD MACHI

Page 51 - ERROR AND STATUS DISPLAYS

8555-M512-O1 Rev A 03/13 49

Page 52 - CARE AND CLEANING

APPENDIX A TOP PLATE MOUNTING CONFIGURATIONS “Type One” Uses Center Mounting Hole “Type Three” Uses Front Mounting Hole

Page 53 - SECURITY SEAL INSTALLATION

“Type Four” “Type Five” Models using “Type Four” Models using “Type Five” 6550 6854DHR 6857DHR 6857KGDHR “Type

Page 54 - PARTS IDENTIFICATION

STATEMENT OF LIMITED WARRANTY Detecto Scale warrants its equipment to be free from defects in material and workmanship as follows: Detecto warrants

Page 55

STATEMENT OF LIMITED WARRANTY Conditions Which Void Limited Warranty This warranty shall not apply to equipment which: A.) Has been tampered with, d

Page 56 - APPENDIX A

8555-M512-O1 Rev A 03/13 6 IV

Page 57

8555-M512-O1 Rev A 03/13 54 Printed in USA Cardinal Scale Mfg. Co. PO Box 151, Webb City, MO 64870Telephone: 417-673-4631 FAX: 417-673-5001 www.de

Page 58

SPECIFICATIONS Power Requirements . . . . . . . . 6 “AA” cell Alkaline, Ni-Cad or NiMH batteries (not included) OR an optional 12 VDC 1.25A wall plug

Page 59 - D268-WARRANTY-DET

SITE PREPARATION REQUIREMENTS The MedVue Model MV1 Medical Weight Analyzer is a precision weight indicating instrument. As with any precision instrum

Page 60 - Cardinal Scale Mfg. Co

INSTALLATION Unpacking Carefully remove the MedVue Model MV1 Medical Weight Analyzer from the shipping carton and inspect it for any damage that may h